Brand Name : drager

Model Number : 6850645

Certification : CE,ISO

Place of Origin : Germany

MOQ : 1pcs

Price : as discussed

Payment Terms : T/T, Western Union,

Supply Ability : 300pcs / week

Delivery Time : 5-7 work days

Packaging Details : 1pcs / bag

COLOR : silver

brand : drager

material : Mixture

warranty : 12 months

name : o2 sensor medical

approval : CE,ISO

Contact Now

Original and new drager medical Oxygen sensor 6850645 for Ventilator 6850645 O2 sensor O2 cell

Sensor Specifications & Compatibility
Catalog Number
Output Signal in Air 24A - 32A (Dual Cathode) no temperature compensation
Response Time T 90% < 12 seconds
Nominal Sensor Lifetime 500 000 % volume Oxygen hours
Electrical Interface Gold plated slip rings
Compatibility: Draeger 6850645 (Dr¤ger)
Babylog, Fabian, Cicero, PM8030
Evita, Oxydig, Anemone, Oxycom
Incubator 8000
M08 ST13
MAX11
Invivo Magnitude
10-103-07
PT14A
Teledyne R13 C-61139
Hudson 5557 (V14)
VTI 101372, 100352
* Note: Order E61720 with reusable cap and "o" rings
Medical Oxygen Sensors (Fuel Cells) general specifications:
Measurement Range: 0-100 % oxygen
Accuracy and Repeatability: < 1 % vol. O2when calibrated at 100 % O2
Zero Ofset: < 0.5 % vol. O2in 100% N2, applied 5 minutes
Linearity Error: < 3 % relative
Cross Interference: < 0.5 % vol. O2response to:
10 % CO2balance N2
80% N2O balance N2
7.5% Halothane balance N2
7.5 % Isofurane balance N2
7.5 % Enfurane balance N2
9% Sevofurane balance N2
20% Desfurane balance N2
Infuence of Humidity: - 0.03 % rel. per % RH at 25C
Infuence of Pressure: proportional to change in oxygen partial
pressure
Infuence of Mechanical Shock: < 1% relative after a fall from 1m
Operating Temperature: 0 to 50C
Temperature Compensation: built-in NTC compensation (see below)
Efect of Temperature between +5 C and +25 C: 3 % relative error
Compensation (steady state): between 0 C and +50 C: 8 % relative error
Operating Humidity: 0-99 % RH non-condensing
Long Term Output Drift: < 1 % vol oxygen per month
typically < - 15 % relative over lifetime
Storage Temperature: -20 to +50 C
Recommended Storage: +5 to +15 C
Recommended Load: 10 kOhms
Warm-Up Time: < 30 minutes, after replacement of sensor
Weight: approximately 28 grams
approximately 43 grams E61912 series
Designed and manufactured according to EN ISO 9001 : 2000 and EN13485 : 2003
specifications are applicable at standard conditions: 1013 hPa, 25C dry ambient air
Oxygen sensor for Draeger OEM 6850645
Envitec 6850645

Ventilator Medical O2 Sensor , Quick React Original New O2 Sensor Medical


Product Tags:

envitec oxygen sensor

      

drager oxygen sensor

      